FDA-Device2022-10-05Class II

MICRO-X ROVER MOBILE X-RAY SYSTEM, MXU-RV19

Micro-X Ltd.
Hazard

Software calibration error with product equip with a Dose Area Product (DAP) meter. This results in the product displaying the incorrect DAP meter values which may impede radiation exposure management decisions

Sold states
Worldwide distribution - U.S. Nationwide distribution in the states of CA, NV, NY, OH, TX and WA and the countries of Australia, France, Israel, Japan, and United Arab Emirates.
Affected count
42 units
Manufactured in
A14 6 MAB Eastern Promenade, N/A, Tonsley, N/A, Australia
Products
MICRO-X ROVER MOBILE X-RAY SYSTEM, MXU-RV19

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1786-2022

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief