FDA-Device2012-06-27Class I
GE Healthcare, Aestiva/5 7900 SmartVent, anesthesia machine. Used in Datex-Ohmeda Aestiva/5 Anesthesia Systems.
GE Healthcare, LLC
Hazard
GE Healthcare has recently become aware of a potential safety issue due to the interlock slide mechanism missing from the 3-vaporizer manifold of your Aestiva 7900 Anesthesia Device. This Potential for two (2) vaporizers delivering agent at the same time. This could manifest as over delivery of a single agent if both vaporizers contain the same agent or as delivery of more than one agent. This m
Sold states
Nationwide Distribution - including the states of AL, FL, GA, MO, and VA.
Affected count
8
Manufactured in
3000 N Grandview Blvd, N/A, Waukesha, WI, United States