FDA-Device2012-06-20Class II
GE Healthcare Revolution XR/d; GE Healthcare, 3000 North Grandview , Waukesha, WI, 53188 Model numbers: 2259988-2 & 2351505. The Revolution XR/d Digital Radiographic Imaging System is indicated for use in generating radiographic images of human anatomy. It is intended to replace radiographic film/screen system in all general-purpose diagnostic procedures. This device is not intended for mammographic applications.
General Electric Med Systems LLC
Hazard
An incident was reported that an operator's finger was pinched between the fixed roller and the gliding tabletop of a Proteus XR/a Table while moving the tabletop with the thumb placed over its edge and fingers underneath
Sold states
Worldwide Distribution.
Affected count
3,796 units
Manufactured in
3000 N Grandview Blvd, Waukesha, WI, United States
Products
GE Healthcare Revolution XR/d; GE Healthcare, 3000 North Grandview , Waukesha, WI, 53188 Model numbers: 2259988-2 & 2351505. The Revolution XR/d Digital Radiographic Imaging System is indicated for use in generating radiographic images of human anatomy. It is intended to replace radiographic film/screen system in all general-purpose diagnostic procedures. This device is not intended for mammographic applications.
Official notice