FDA-Device2017-04-19Class II
VIDAS Estradiol II, Ref 30431-01, 60 tests The VIDAS Estradiol II (E2 II) assay is intended for use on the instruments of the VIDAS family (VITEK ImmunoDiagnostic Assay System) as an automated quantitative enzyme-linked fluorescent immunoassay (ELFA) to the determination of total estradiol concentration in human serum or plasma (heparin).
BioMerieux SA
Hazard
Fulvestrant cross reacts with the VIDAS Estradiol II immunoassay and may cause falsely elevated estradiol results in patients treated with Fulvestrant.
Sold states
Nationwide Distribution including AZ, CA, CO, FL, GA, IN, IA, LA, MI, MS, NY, NC, OH, OK, OR, PA, TX, WA
Affected count
834 kits
Manufactured in
Chemin De L'Orme, N/A, Marcy L'Etoile, N/A, France
Products
VIDAS Estradiol II, Ref 30431-01, 60 tests The VIDAS Estradiol II (E2 II) assay is intended for use on the instruments of the VIDAS family (VITEK ImmunoDiagnostic Assay System) as an automated quantitative enzyme-linked fluorescent immunoassay (ELFA) to the determination of total estradiol concentration in human serum or plasma (heparin).
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1801-2017Don't want to check this manually?