FDA-Device2020-05-06Class II

Obalon Balloon System with Model 4300 Touch Dispenser Touchscreen; Software version 02.00.00.0076 - Product Usage: The System is intended to be used as an adjunct to a moderate intensity diet and behavior modification program. All balloons must be removed 6 months after the first balloon is placed. The Obalon Balloons are administered using the Navigation System.

Obalon Therapeutics Inc
Hazard

During initial set-up and/or replacement of the dispenser batteries, the touch dispenser touchscreen can inadvertently lose calibration. This issue can also occur if the touchscreen is pressed when the device is powered on in preparation for a balloon administration. As a result, the touchscreen can become unresponsive and balloon inflation cannot be initiated.

Sold states
US Nationwide distribution including the states of VA, NY, GA, TX, TN, OR, FL, LA, CT, and NJ.
Affected count
45 units
Manufactured in
5421 Avenida Encinas Ste F, N/A, Carlsbad, CA, United States
Products
Obalon Balloon System with Model 4300 Touch Dispenser Touchscreen; Software version 02.00.00.0076 - Product Usage: The System is intended to be used as an adjunct to a moderate intensity diet and behavior modification program. All balloons must be removed 6 months after the first balloon is placed. The Obalon Balloons are administered using the Navigation System.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1805-2020