Philips Azurion 7 M20 -XperGuide Software hosted in Interventional Workspot 1.5 when used with the Azurion R2.0 system Interventional fluoroscopic x-ray system System code: 722079

Philips North America, LLC
Hazard

When a user acquires XperCT scan on an Azurion 2.0 system, enters the XperGuide guidance step and moves the L-arm away from the initial scan position before starting the live guidance, a warning message directs the user to move the L-arm stand back to the initial XperCT scan position. Although the software generates this message, it does not prevent the use of live guidance if the L-arm stand is not repositioned. Using live guidance with a mispositioned L-arm can result in the display of an incorrect overlay and needle path.

Sold states
Worldwide distribution. US states of FL, ID, IL, IN, MA, and OH, Canada, Australia, Austria, Belgium, Germany, and Netherlands New Zealand, South Africa, Switzerland, Thailand
Affected count
22 units (6 US and 16 OUS); Expanded 3units; Total: 25
Manufactured in
3000 Minuteman Rd, N/A, Andover, MA, United States
Products
Philips Azurion 7 M20 -XperGuide Software hosted in Interventional Workspot 1.5 when used with the Azurion R2.0 system Interventional fluoroscopic x-ray system System code: 722079

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1809-2020