FDA-Device2024-05-22Class II

LASER Application Instrument, 23 cm, REF: 8694; LEUNIG-GREVERS LASER Application, REF: 461000

Karl Storz Endoscopy
Hazard

Laser surgical instruments have no specific evidence of a validated reprocessing method, therefore these instruments lack sterility assurance, so once the instruments have gone through reprocessing after use and then are used again, there is an increased risk of the patient being exposed to infection.

Sold states
Worldwide distribution - US Nationwide in the states of OH, CA, FL, MA, TX, KY, WI, MI, MN, AR, NE, IL, NY, HI and the country of Argentina.
Affected count
85
Manufactured in
2151 E Grand Ave, N/A, El Segundo, CA, United States
Products
LASER Application Instrument, 23 cm, REF: 8694; LEUNIG-GREVERS LASER Application, REF: 461000

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1809-2024