FDA-Device2022-10-05Class II

Dimension Magnesium (MG) Flex reagent cartridge Siemens Material Number (SMN)/REF (Catalog Number):10444963/DF57

Siemens Healthcare Diagnostics, Inc.
Hazard

Dimension Magnesium (MG) Flex reagent cartridge lots may exhibit imprecision for Quality Control (QC) and produce Abnormal Reaction flags. The issue is not always detected by QC and erroneous results may be produced in the absence of an Abnormal Reaction flag. Based on customer data, imprecision leading to erroneous, unflagged patient results with a bias of -15 to -59% may occur

Sold states
Nationwided Foreign: Afghanistan¿¿¿¿ Albania¿¿¿¿¿¿¿¿ Austria¿¿¿¿¿¿¿¿ Bahrain¿¿¿¿¿¿¿¿ Bosnia¿Herzeg.¿ Bulgaria¿¿¿¿¿¿¿ Canada Croatia¿¿¿¿¿¿¿¿ Czech¿Republic¿ Egypt¿¿¿¿¿¿¿¿¿¿ France¿¿¿¿¿¿¿¿¿ Georgia¿¿¿¿¿¿¿¿ Germany¿¿¿¿¿¿¿¿ Greece¿¿¿¿¿¿¿¿¿ Hungary¿¿¿¿¿¿¿¿ Israel¿¿¿¿¿¿¿¿¿ Italy¿¿¿¿¿¿¿¿¿¿ Jordan¿¿¿¿¿¿¿¿¿ Kazakhstan¿¿¿¿¿ Kenya¿¿¿¿¿¿¿¿¿¿ Kuwait¿¿¿¿¿¿¿¿¿ Latvia¿¿¿¿¿¿¿¿¿ Lebanon¿¿¿¿¿¿¿¿ Macedonia¿¿¿¿¿¿ Mali¿¿¿¿¿¿¿¿¿¿¿ Malta¿¿¿¿¿¿¿¿¿¿ Montenegro¿¿¿¿¿ Morocco¿¿¿¿¿¿¿¿ Nepal¿¿¿¿¿¿¿¿¿¿ Netherlands¿¿¿¿ Pakistan¿¿¿¿¿¿¿ Poland¿¿¿¿¿¿¿¿¿ Portugal¿¿¿¿¿¿¿ Qatar¿¿¿¿¿¿¿¿¿¿ Romania¿¿¿¿¿¿¿¿ Russian¿Fed.¿¿¿ Saudi¿Arabia¿¿¿ Senegal¿¿¿¿¿¿¿¿ Serbia¿¿¿¿¿¿¿¿¿ Slovakia¿¿¿¿¿¿¿ Slovenia¿¿¿¿¿¿¿ Spain¿¿¿¿¿¿¿¿¿¿ Sudan¿¿¿¿¿¿¿¿¿¿ Switzerland¿¿¿¿ Turkey¿¿¿¿¿¿¿¿¿ U.A.E.¿¿¿¿¿¿¿¿¿ Uganda¿¿¿¿¿¿¿¿¿ Ukraine¿¿¿¿¿¿¿¿ United¿Kingdom¿ Uzbekistan¿¿¿¿¿
Affected count
26851 units
Manufactured in
500 Gbc Dr, N/A, Newark, DE, United States
Products
Dimension Magnesium (MG) Flex reagent cartridge Siemens Material Number (SMN)/REF (Catalog Number):10444963/DF57

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1816-2022

Don't want to check this manually?