FDA-Device2017-04-19Class II
Nordson Micromedics FibriJet Endoscopic Catheter Introducer, Catalog Number SA-3445. Sterilized using Ethylene Oxide
Micromedics, Inc.
Hazard
The sterile packaging may contain small channels in the seal of the pouch, If the channels are present, the package integrity may be compromised which may affect sterility assurance.
Sold states
AL, CA, FL, MS, NC, NJ, NY, TN, and TX. Outside the US: Israel
Affected count
273 units
Manufactured in
1270 Eagan Industrial Rd Ste 120, N/A, Saint Paul, MN, United States
Products
Nordson Micromedics FibriJet Endoscopic Catheter Introducer, Catalog Number SA-3445. Sterilized using Ethylene Oxide