FDA-Device2024-05-22Class II

HeartSine samaritan PAD (Public Access Defibrillator), Type: SAM 350P, SAM 360P, SAM 450P

HeartSine Technologies Ltd
Hazard

Automated external defibrillators have a manufacturing issue that could prevent the device from delivering instructional voice prompts to the user during use of the device. Visual instructional icons will still be present, but this issue could potentially lead to no therapy or a delay in therapy.

Sold states
US Nationwide distribution including in the states of AK, AL, AR, AS, AZ, CA, CO, CT, DC, DE, FL, GA, GC, HI, IA, ID, IL, IN, KS, KY, LA, MA, MB, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV, WY.
Affected count
183,248
Manufactured in
207 Airport Road West, N/A, Belfast, N/A, United Kingdom
Products
HeartSine samaritan PAD (Public Access Defibrillator), Type: SAM 350P, SAM 360P, SAM 450P

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1820-2024

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief