FDA-Device2017-04-26Class II

UNIQ: UNIQ FD10 722026; UNIQ FD10/10 722027; UNIQ FD20/10 biplane 722029; UNIQ FD10 OR Table 722033; UNIQ FD10/10 OR Table 722034; UNIQ FD20/10 OR Table 722036; UNIQ FD20/20 722038; UNIQ FD20/20 biplane OR Table 722039

Philips Electronics North America Corporation
Hazard

Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.

Sold states
Nationwide, worldwide including Canada
Affected count
6992 (total for all devices)
Manufactured in
3000 Minuteman Rd, N/A, Andover, MA, United States
Products
UNIQ: UNIQ FD10 722026; UNIQ FD10/10 722027; UNIQ FD20/10 biplane 722029; UNIQ FD10 OR Table 722033; UNIQ FD10/10 OR Table 722034; UNIQ FD20/10 OR Table 722036; UNIQ FD20/20 722038; UNIQ FD20/20 biplane OR Table 722039

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1821-2017