FDA-Device2024-05-22Class II
The Sophysa Pressio¿ Ventricular Intracranial Pressure Monitoring Kit monitoring is intended for monitoring intracranial ventricular pressure and temperature with tunneling (PSO-VTT) indicated for use in intraventricular pressure monitoring and cerebrospinal fluid drainage application. Model/Catalog Number: PSO-VTT
Sophysa S.A.
Hazard
An increase in claims related to leakage of cerebrospinal fluid (CSF) from the proximal end of the catheter. In the event of a leak going undetected during implantation, it could pose an infectious risk to the patient. The cause of the leaks is indicated to be manufacturing error involving perforation of the inner channel of the catheter during stylet insertion.
Sold states
Worldwide - US Nationwide distribution in the states of AZ, NY, TX and the countries of ARGENTINA, CHINA, CZECH REPUBLIC, FRANCE, GREECE, HUNGARY, JAPAN, PAKISTAN, Portugal, ROMANIA, SOUTH AFRICA, SWEDEN, VIETNAM.
Affected count
3175 units- 41 units US; 3134 units OUS
Manufactured in
5 Rue Guy Moquet, N/A, Orsay, N/A, France
Products
The Sophysa Pressio¿ Ventricular Intracranial Pressure Monitoring Kit monitoring is intended for monitoring intracranial ventricular pressure and temperature with tunneling (PSO-VTT) indicated for use in intraventricular pressure monitoring and cerebrospinal fluid drainage application. Model/Catalog Number: PSO-VTT
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1826-2024Don't want to check this manually?