FDA-Device2017-04-26Class II

Proteus 235, Proton Therapy System

Ion Beam Applications S.A.
Hazard

In specific circumstances, the user is able to restart an aborted irradiation without having to select the correct partial irradiation data.

Sold states
Distribution in US (to Louisiana), and France.
Affected count
2 worldwide, 1 in U.S.
Manufactured in
Chemin du Cyclotron, 3, N/A, Louvain La Neuve, N/A, Belgium
Products
Proteus 235, Proton Therapy System

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1830-2017