FDA-Device2013-08-07Class II
Exactech Articulated Driver A nonpowered hand-held device intended for medical purposes to manipulate tissue, or for use with other devices in orthopedic surgery.
Exactech, Inc.
Hazard
When torque is applied to the Articulated Driver at the maximum angle, the cross pin may shift out of position, potentially leading to disassembly of the instrument.
Sold states
US Distribution including the states of AR, CA, FL, NJ, NY, OK, and TX
Affected count
57
Manufactured in
2320 NW 66th Ct, N/A, Gainesville, FL, United States
Products
Exactech Articulated Driver A nonpowered hand-held device intended for medical purposes to manipulate tissue, or for use with other devices in orthopedic surgery.
Official notice