FDA-Device2013-08-14Class I

Alaris PC unit, model 8015, with version 9.12-Keyboard Processor and logic board assembly part number TC10006584. The Alaris PC Unit is the main user interface unit and power supply of the Alaris System, a modular system to be used with Alaris System modules intended for use in todays growing professional healthcare environment for facilities that utilize infusion and/or monitoring devices.

CareFusion 303, Inc.
Hazard

CareFusion is recalling the Alaris PC units model 8015 (PC unit), version 9.12, because it is operating at an incorrect voltage and it could potentially experience a loss of communication between the PC Unit main Processor and Keyboard Processor.

Sold states
Worldwide distribution: US (nationwide) and countries of: Australia, Canada and Europe.
Affected count
26,564 total units (26,220 units for US)
Manufactured in
10020 Pacific Mesa Blvd, N/A, San Diego, CA, United States
Products
Alaris PC unit, model 8015, with version 9.12-Keyboard Processor and logic board assembly part number TC10006584. The Alaris PC Unit is the main user interface unit and power supply of the Alaris System, a modular system to be used with Alaris System modules intended for use in todays growing professional healthcare environment for facilities that utilize infusion and/or monitoring devices.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1835-2013