FDA-Device2012-06-27Class II
SARNS FLEXIBLE ARTERIAL CANNULA: 8.0 MM (24 FR) WITH 3/8" CONNECTOR,SUTURE RING, 9.5" (24 CM) LONG. The Sarns High Flow Aortic Arch Cannula is indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery.
Terumo Cardiovascular Systems Corporation
Hazard
During production of one lot of the Sarns High-Flow Aortic Arch Cannula, Terumo Cardiovascular Systems (Terumo CVS) found a foreign substance on the outer surface of some cannulae connectors. Terumo CVS' preliminary investigation found that the substance can be dislodged from the cannula surface, the substance was likely deposited during the molding process, but the exact composition of the parti
Sold states
Worldwide Distribution-USA (nationwide) including the states of AL , AR , AZ, CA , CO , CT , FL , GA , HI , IA , IL , IN , KS , KY , LA , MA , MD , ME , MI , MI , MN , MO , MT , NC , ND , NE , NH , NJ , NM , NV , NY , OH , OK , OR , PA , RI , SC , SD , TN , TX , UT , VA , WA , WI, and WY and the countries of Australia, Belgium, Brazil, Canada, Chile, Germany, Honduras, Hong Kong, India, Japan, Mauritius, Nicaragua, Philippines, Saudi Arabia, Singapore, South Africa, Taiwan, Thailand, Trinidad and Tobago, United Arab Emirates (UAE), and Vietnam.
Affected count
133,793 (each)
Manufactured in
6200 Jackson Road, N/A, Ann Arbor, MI, United States
Products
SARNS FLEXIBLE ARTERIAL CANNULA: 8.0 MM (24 FR) WITH 3/8" CONNECTOR,SUTURE RING, 9.5" (24 CM) LONG. The Sarns High Flow Aortic Arch Cannula is indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery.
Official notice