FDA-Device2020-05-13Class II

Transducer S7-3t UDI (01)00884838061439 REF 989605406772 Can be used with: CX50, EPIQ 7, Affiniti 50 and Affiniti 70 Ultrasound systems - Product Usage: Diagnostic ultrasound imaging or fluid flow analysis of the human body - particularly transesophageal echocardiography (TEE) studies

Philips Ultrasound Inc
Hazard

The firm has discovered an issue associated with some Transesophageal (TEE) transducers that were not properly programmed during manufacturing. As a result, the system is unable to correctly measure the temperature on the distal tip of the transducer. This issue may result in affected transducers not initiating auto-cool during use which could result in patient injury

Sold states
Worldwide distribution - US Nationwide distribution.
Affected count
N/A
Manufactured in
22100 Bothell Everett Hwy, N/A, Bothell, WA, United States
Products
Transducer S7-3t UDI (01)00884838061439 REF 989605406772 Can be used with: CX50, EPIQ 7, Affiniti 50 and Affiniti 70 Ultrasound systems - Product Usage: Diagnostic ultrasound imaging or fluid flow analysis of the human body - particularly transesophageal echocardiography (TEE) studies

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1841-2020