FDA-Device2022-10-05Class II
MEDLINE CLEARPRO DOUBLE SWIVEL CLOSED SUCTION CATHETER, flat peel pouch EA, 10 EA per Inner Carton (IC), 20 EA/2 Inner cartons per Case (CS). Labeled as follows: a. ET LENGTH DOUBLE SWIVEL, 12 Fr, Item Number/REF: DYNCPDS12; b. ET LENGTH DOUBLE SWIVEL, 14 Fr, Item Number/REF: DYNCPDS14; c. ET LENGTH DOUBLE SWIVEL, 16 Fr, Item Number/REF: DYNCPDS16
MEDLINE INDUSTRIES, LP - Northfield
Hazard
Potential for the suction catheter to come apart from the green connector near the thumb valve. The connector is coming apart from the assembled device when pulling the catheter out of the artificial airway in the patient suctioning process.
Sold states
Nationwide distribution to AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and Puerto Rico. International distribution to Canada, Chile, Colombia, Panama, Qatar, United Arab Emirates
Affected count
65,320 units
Manufactured in
3 Lakes Dr, N/A, Northfield, IL, United States
Products
MEDLINE CLEARPRO DOUBLE SWIVEL CLOSED SUCTION CATHETER, flat peel pouch EA, 10 EA per Inner Carton (IC), 20 EA/2 Inner cartons per Case (CS). Labeled as follows: a. ET LENGTH DOUBLE SWIVEL, 12 Fr, Item Number/REF: DYNCPDS12; b. ET LENGTH DOUBLE SWIVEL, 14 Fr, Item Number/REF: DYNCPDS14; c. ET LENGTH DOUBLE SWIVEL, 16 Fr, Item Number/REF: DYNCPDS16
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1855-2022Don't want to check this manually?