FDA-Device2021-06-30Class I

GENOSYL DS; Nitric Oxide Delivery System Combination Product NDA202860

Vero Biotech, LLC
Hazard

Reports received of NO measured below desired dose during transition between primary console and backup console since Software Version 2.2.3 was uploaded to consoles in the field.

Sold states
Distributed nationwide to TX, MD, VA, GA, FL, KS, LA, MN, and NC
Affected count
465 units
Manufactured in
387 Technology Cir Nw Ste 125, N/A, Atlanta, GA, United States
Products
GENOSYL DS; Nitric Oxide Delivery System Combination Product NDA202860

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1863-2021