FDA-Device2016-06-08Class II
Torque Limiting Handle: PRECIMED, T-Handle Torque, Non Sterile. Handle: Customer Number/Model: MCL120014 T14000, TLT911-50-S20 TLT911-50-S20, TLT911-15-S01 TLT911-15-S01, TLTS0007-20-S19 TLTS0007-20-S19, TLT911-35-S01 TLT911-35-S01.
Greatbatch Medical
Hazard
Greatbatch Medical has initiated a global Field Recall for Torque Limiting Devices. Greatbatch identified through laboratory testing that the required sterility assurance level (SAL) of 10-6 is not achieved on torque limiting devices when subjected to steam sterilization cycles identified in the Instruction For Use (IFU) provided with the device.
Sold states
US: IN, MI, TX. OUS: GERMANY, FRANCE, ITALY, SWITZERLAND.
Affected count
2
Manufactured in
2300 Berkshire Ln N, N/A, Plymouth, MN, United States
Products
Torque Limiting Handle: PRECIMED, T-Handle Torque, Non Sterile. Handle: Customer Number/Model: MCL120014 T14000, TLT911-50-S20 TLT911-50-S20, TLT911-15-S01 TLT911-15-S01, TLTS0007-20-S19 TLTS0007-20-S19, TLT911-35-S01 TLT911-35-S01.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1882-2016Don't want to check this manually?