FDA-Device2016-06-08Class II

Torque Limiting Screwdriver: Greatbatch Medical, Torque Limiting Driver, Non Sterile, Rx Only. PRECIMED, Non Sterile. Zimmer, Non Sterile, Rx Only. Screwdriver: Customer Number/Model: 804-03-038 T11623, T14001 T14001, MTO35173507 T14378, MTO375173506 T14379, MTO35173505 T14380, MTO35175303 T8127.

Greatbatch Medical
Hazard

Greatbatch Medical has initiated a global Field Recall for Torque Limiting Devices. Greatbatch identified through laboratory testing that the required sterility assurance level (SAL) of 10-6 is not achieved on torque limiting devices when subjected to steam sterilization cycles identified in the Instruction For Use (IFU) provided with the device.

Sold states
US: IN, MI, TX. OUS: GERMANY, FRANCE, ITALY, SWITZERLAND.
Affected count
1110
Manufactured in
2300 Berkshire Ln N, N/A, Plymouth, MN, United States
Products
Torque Limiting Screwdriver: Greatbatch Medical, Torque Limiting Driver, Non Sterile, Rx Only. PRECIMED, Non Sterile. Zimmer, Non Sterile, Rx Only. Screwdriver: Customer Number/Model: 804-03-038 T11623, T14001 T14001, MTO35173507 T14378, MTO375173506 T14379, MTO35173505 T14380, MTO35175303 T8127.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1883-2016

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief