FDA-Device2017-05-03Class II

SmartSite Add-On Bag Access Device, Model No. 10013365

CareFusion 303, Inc.
Hazard

CareFusion is recalling the SmartSite Add-On Bag Device due to reports of separation and/or leakages between the spike port and the drip chamber spike. A separation or leak can cause a delay of infusion, an interruption of infusion, exposure to medication or hazardous infusates, or underinfusion.

Sold states
US Distribution
Affected count
384,923 units
Manufactured in
10020 Pacific Mesa Blvd, N/A, San Diego, CA, United States
Products
SmartSite Add-On Bag Access Device, Model No. 10013365

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1883-2017