FDA-Device2016-06-08Class II
Torque Limiting Adapter: Torque Limiting Attachment, Non Sterile, Distributed by Zimmer. PRECIMED, Torque Limiting Adaptor, Non Sterile.. Customer Number/Model: 00-2360-080-00 T10232, T116245 T116245, TLA025-25 TLA025-25, 00-2360-080-05 T10233, MAC21003 T116430, TLA025-30 TLA025-30, MAC21001 T112972, MAC21001 T13621, TLA025-20 TLA025-20.
Greatbatch Medical
Hazard
Greatbatch Medical has initiated a global Field Recall for Torque Limiting Devices. Greatbatch identified through laboratory testing that the required sterility assurance level (SAL) of 10-6 is not achieved on torque limiting devices when subjected to steam sterilization cycles identified in the Instruction For Use (IFU) provided with the device.
Sold states
US: IN, MI, TX. OUS: GERMANY, FRANCE, ITALY, SWITZERLAND.
Affected count
5952
Manufactured in
2300 Berkshire Ln N, N/A, Plymouth, MN, United States
Products
Torque Limiting Adapter: Torque Limiting Attachment, Non Sterile, Distributed by Zimmer. PRECIMED, Torque Limiting Adaptor, Non Sterile.. Customer Number/Model: 00-2360-080-00 T10232, T116245 T116245, TLA025-25 TLA025-25, 00-2360-080-05 T10233, MAC21003 T116430, TLA025-30 TLA025-30, MAC21001 T112972, MAC21001 T13621, TLA025-20 TLA025-20.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1884-2016Don't want to check this manually?