FDA-Device2020-05-13Class II
Inpeco FlexLab Centrifuge Module tube loading algorithm, Inpeco SA P/N: FLX-202, with Firmware Versions: CM_1-7-0.H86 or xCMd_2-5-0.elf and CM_1-7-1.H86 or xCMd_2-5-1-849fad9.elf
Inpeco S.A.
Hazard
The FlexLab Centrifuge Module loading algorithm may lead to unbalanced loads which could damage the centrifuge.
Sold states
Worldwide distribution including the following states: Ohio, New York, Virginia, Pennsylvania, Texas, Delaware, California, Florida, Kentucky, Tennessee, Maryland, Rhode Island, Alabama, Massachusetts, Connecticut, Minnesota, Alabama, Montana, Louisiana, New Jersey, Virginia, New Hampshire, Michigan, North Carolina, Washington, Iowa, New Mexico, Nebraska, and Missouri
Affected count
65 units
Manufactured in
Via San Gottardo 10, N/A, Lugano, N/A, Switzerland
Products
Inpeco FlexLab Centrifuge Module tube loading algorithm, Inpeco SA P/N: FLX-202, with Firmware Versions: CM_1-7-0.H86 or xCMd_2-5-0.elf and CM_1-7-1.H86 or xCMd_2-5-1-849fad9.elf