FDA-Device2018-05-23Class II

Philips Volcano CORE M2 Vascular System, Part # 400-0100.17, Rx Only. Used for the qualitative and quantitative evaluation of vascular morphology in the coronary arteries and vessels of the peripheral vasculature.

Volcano Corporation
Hazard

The incorrect resistor has been used in affected CORE M2 Imaging Systems, which causes noise and results in an overly bright IVUS Image.

Sold states
Worldwide Distribution -- U.S., to the states of California, Connecticut, Tennessee, North Carolina, Maryland, and Oregon; and, the country of Australia.
Affected count
8 devices
Manufactured in
2870 Kilgore Rd, Rancho Cordova, CA, United States
Products
Philips Volcano CORE M2 Vascular System, Part # 400-0100.17, Rx Only. Used for the qualitative and quantitative evaluation of vascular morphology in the coronary arteries and vessels of the peripheral vasculature.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1891-2018