FDA-Device2024-06-05Class II
Philips Allura Xper FD20 Biplane system with an ADN7NT patient table. Model Number: 722013, Software Version Number 8.1.100.
Hazard
When Software upgrade 8.1.100 is installed on an Allura Xper with an AD7NT (non tilt) table, the pivot brake does not engage. This issue may cause unexpected movements of the table.
Sold states
Domestic: OK International: India, Spain, U.A.E.
Affected count
1 unit (OUS)
Manufactured in
Veenpluis 4-6, Best, Netherlands
Products
Philips Allura Xper FD20 Biplane system with an ADN7NT patient table. Model Number: 722013, Software Version Number 8.1.100.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1897-2024More PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls
- FDA-Device2026-06-03Philips Azurion imaging systems recalled for hard drive failure risk affecting imaging functionality
- FDA-Device2026-06-03Philips Allura medical imaging systems recalled for hard drive failure risk
- FDA-Device2026-05-27Philips Azurion X-ray systems recalled for cooling hose degradation risk
- FDA-Device2026-05-27Philips Integris-Allura X-ray systems recalled for cooling unit oil leakage risk
- FDA-Device2026-05-27Philips Azurion X-ray systems recalled for impaired table movement due to control wear