FDA-Device2021-06-30Class II
Access SARS CoV-2 lgG II Reagent, REF: C69057, IVD CE UDI: (01)15099590742744
Beckman Coulter Inc.
Hazard
SARS-CoV-2 lgG II numerical results may potentially be multiplied by a factor of 1000 on systems running with assay protocol file (APF) and access assay file (AAF) versions resulting in falsely elevated numerical values.
Sold states
US: Alabama, Alaska, Arizona, Arkansas, California, Florida, Georgia, Illinois, Iowa, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Mississippi, Missouri, New Hampshire, New Jersey, New York, North Dakota, Ohio, Oklahoma, Pennsylvania, Puerto Rico, Tennessee, Texas, Virginia, West Virginia, Wisconsin, Wyoming, OUS: Andorra, Austria, Croatia, Czech Republic, France, Germany, Gibraltar, Greece, Hong Kong, Hungary, Israel, Italy, Japan, Jordan, Poland, Portugal, San Marino, Slovakia, Slovenia, Spain, United Arab Emirates
Affected count
13,337 units
Manufactured in
250 S Kraemer Blvd, N/A, Brea, CA, United States
Products
Access SARS CoV-2 lgG II Reagent, REF: C69057, IVD CE UDI: (01)15099590742744
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1905-2021Don't want to check this manually?