FDA-Device2025-06-18Class I

Newport HT70 Ventilator, REF: HT70M-JP-NA, HT70M-CN-NA, HT70M-ES-EU, HT70M-PT-BR, HT70M-SY-EU, HT70M-WW-EU, HT70M-WW-NA, HT70-SY-EU. Newport HT70Plus Ventilator, REF: HT70PM-ES-EU, HT70PM-ES-NA, HT70PM-JP-NA, HT70PM-PT-BR, HT70PM-SY-AS, DLHT70PM-WW-NA, HT70PM-SY-NA, HT70PM-SY-UK, HT70PM-WW-EU, HT70PM-WW-UK, HT70PM-SY-EU, HT70PM-WWNA. HT70-2 CONTROL BOARD ROHS X1, REF: GR105796. HT70P CONTROL BOARD ROHS X1, REF: GR105795. HT70 Conversion Kit, REF: GRFSKHT70M-2.

Covidien
Hazard

Ventilator Printed Circuit Board Assembly may have two separate capacitors that may fail, which may result in the ventilator either shutting down during use, thus necessitating use of an alternate form of ventilation, or the shutdown alert alarm fails to alarm effectively during shut down, which may result in respiratory failure, hypoventilation, low oxygen saturation, hypoxia, treatment delay.

Sold states
Worldwide - US Nationwide distribution including in the states of MA, CT, KS, OH, PR, NH, GA, WI, MS, VA, CO, TX, MN, AL, IA, NY, UT, NM, KY, RI, SD, PA, IA, CA, MO, HI, MI, IL, MT, NJ, TN, SC, FL, IN, AZ, ID, LA, WA, NC, MD, WY, NE, NV, ND, ME, AK, DE, OK, OR and the countries of Japan, Canada, Singapore, Lithuania, Kazakhstan, Slovenia, Colombia, China, Qatar, United Arab Emirates, Greece, Saudi Arabia, Nigeria, Romania, Netherlands, Lebanon, Oman, Jordan, Malawi, India, Croatia, Portugal, Turkey, Brazil, Pakistan, Trinidad and Tobago, Slovakia, Chile, Egypt, Switzerland, South Africa, Thailand, Bangladesh, Myanmar, Peru, Uruguay, Hungary, Argentina, Iran, Islamic Republic Of, Costa Rica, Vietnam, Australia, Guatemala, Mexico, Guyana, Jamaica, Dominican Republic, Bahamas, Bosnia and Herzegovina, South Korea, Tunisia, Ukraine, Uzbekistan, Indonesia, Burkina Faso, Italy, Panama, Sri Lanka, Malaysia, Namibia, Swaziland, Spain, Philippines, Bhutan, Mauritius, Poland, France, Cambodia, Czechia, Israel, Norway, Russian Federation, United Kingdom, Canary Islands, Kenya, Czech Republic, Croatia , Estonia, Maldova, Palestine.
Affected count
4842
Manufactured in
2101 Faraday Ave, Carlsbad, CA, United States
Products
Newport HT70 Ventilator, REF: HT70M-JP-NA, HT70M-CN-NA, HT70M-ES-EU, HT70M-PT-BR, HT70M-SY-EU, HT70M-WW-EU, HT70M-WW-NA, HT70-SY-EU. Newport HT70Plus Ventilator, REF: HT70PM-ES-EU, HT70PM-ES-NA, HT70PM-JP-NA, HT70PM-PT-BR, HT70PM-SY-AS, DLHT70PM-WW-NA, HT70PM-SY-NA, HT70PM-SY-UK, HT70PM-WW-EU, HT70PM-WW-UK, HT70PM-SY-EU, HT70PM-WWNA. HT70-2 CONTROL BOARD ROHS X1, REF: GR105796. HT70P CONTROL BOARD ROHS X1, REF: GR105795. HT70 Conversion Kit, REF: GRFSKHT70M-2.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1915-2025