FDA-Device2015-07-08Class II

Cholesterol Gen.2. The cassette COBAS Integra Cholesterol Gen.2 (CHOL2) contains an in vitro diagnostic reagent system intended for use on COBAS Integra 800 and 400 plus systems, cobas c111, and cobas c systems for the quantitative determination of total cholesterol in serum and plasma.

Roche Diagnostics Operations, Inc.
Hazard

Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.

Sold states
US (nationwide) including PR.
Affected count
6,255 analytical units distributed
Manufactured in
9115 Hague Rd, Indianapolis, IN, United States
Products
Cholesterol Gen.2. The cassette COBAS Integra Cholesterol Gen.2 (CHOL2) contains an in vitro diagnostic reagent system intended for use on COBAS Integra 800 and 400 plus systems, cobas c111, and cobas c systems for the quantitative determination of total cholesterol in serum and plasma.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1921-2015