FDA-Device2020-05-13Class II

Philips Brilliance iCT Computed Tomography X-Ray System; Model # 728306; UDI #s (01)00884838059474(21)100783, (01)00884838059474(21)100848

Philips North America, LLC
Hazard

The bolt affixing the cooling unit to the rotor was not tightened to specification on the Philips Brilliance iCT systems.

Sold states
Worldwide distribution including one unit distributed to Iowa.
Affected count
19
Manufactured in
3000 Minuteman Rd, N/A, Andover, MA, United States
Products
Philips Brilliance iCT Computed Tomography X-Ray System; Model # 728306; UDI #s (01)00884838059474(21)100783, (01)00884838059474(21)100848

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1921-2020