FDA-Device2018-05-30Class II
Prismaflex 7.20 US, Product Code 955542 Intended Use: The Prismaflex Control Unit is intended for: " Continuous Renal Replacement Therapy (CRRT) for patients weighing 20 kilograms or more with acute renal failure and/or fluid overload. " Therapeutic Plasma Exchange (TPE) therapy for patients weighing 20 kilograms or more with diseases where removal of plasma components is indicated. " Continuous Renal Replacement Therapy (CRRT) in conjunction with the MARS system to conduct MARS treatments f
Baxter Healthcare Corporation
Hazard
Baxter Healthcare will be installing new firmware on all Prismaflex control units to address the potential for a small number of these units to exhibit a failure mode with the pump module electronics. The failure mode may result in a Voltage Out of Range malfunction alarm, which causes the device to enter a safe state and become inoperable until it is serviced. Baxter will be releasing new firmware that will prevent the malfunction from occurring.
Sold states
Worldwide distribution: United States (including Puerto Rico), and countries of: Canada, Australia, New Zealand, China, Bangladesh, Japan, Malaysia, Myanmar, Sri Lanka, India, Singapore, Vietnam, Hong Kong, Korea, Taiwan, Indonesia, Bruni, Philippines, Thailand, Argentina, Chile, Costa Rica, Dominican Republic, Haiti, Barbados, Bermuda, Venezuela, Cuba, Mexico, Colombia, Brazil and EMEA.
Affected count
231 units
Manufactured in
1 Baxter Pkwy, N/A, Deerfield, IL, United States
Products
Prismaflex 7.20 US, Product Code 955542 Intended Use: The Prismaflex Control Unit is intended for: " Continuous Renal Replacement Therapy (CRRT) for patients weighing 20 kilograms or more with acute renal failure and/or fluid overload. " Therapeutic Plasma Exchange (TPE) therapy for patients weighing 20 kilograms or more with diseases where removal of plasma components is indicated. " Continuous Renal Replacement Therapy (CRRT) in conjunction with the MARS system to conduct MARS treatments for patients weighing 20 kilograms or more. All treatments administered via the Prismaflex control unit must be prescribed by a physician.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1926-2018Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief