FDA-Device2018-05-30Class II
Edwards Lifesciences' IntraClude Intra-Aortic Occlusion Device. Packaged in a sterile barrier system. Model Number: ICF100. Designed to occlude the ascending aorta.
Edwards Lifesciences, LLC
Hazard
Possible leak at distal open end of the strain relief, which causes medical staff to continue adding inflation fluid to maintain balloon pressure for adequate occlusion of the ascending aorta.
Sold states
Worldwide Distribution -- USA, to the states of AL, AZ, CA, CO, CT, FL, GA, IL, IN, KY, MA, MI, MS, MO, NV, NJ, NY, NC, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, and WV; and, the countries of to Australia, Austria, Belgium, Canada, Colombia, Czech Republic, France, Germany, Israel, Italy, Luxembourg, Netherlands, Poland, Romania, Spain, Sweden, Switzerland, United Arab Emirates, and United Kingdom.
Affected count
3,052 units total
Manufactured in
12050 S Lone Peak Pkwy, N/A, Draper, UT, United States
Products
Edwards Lifesciences' IntraClude Intra-Aortic Occlusion Device. Packaged in a sterile barrier system. Model Number: ICF100. Designed to occlude the ascending aorta.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1929-2018Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.