FDA-Device2013-08-21Class II
CELLTRACKS ANALYZER II (Software Version 2.5.1 & Below. Product Usage: The CELLTRACKS AUTOPREP System is a general purpose laboratory instrument used with immunomagnetic reagents that capture and enrich target cells, and labeling reagents that differentiate cells in whole blood. Cell analyzers such as the CellTracks Analyzer II, CellSpotter System, flow cytometers or microscopes may be used for cell identification and enumeration. The system is fit for in vitro diagnostic use. In the CellT
Veridex, LLC
Hazard
Veridex, LLC has received a report that a patient's demographics were incorrectly associated with another patient's identification (PID) number. The purpose of this notification is to inform you of the issue and provide instructions to help prevent an occurrence on your CELLTRACKS AUTOPREP System and CELLTRACKS ANALYZER II.
Sold states
Worldwide Distribution - USA Nationwide and the countries of Australia, Canada, China, India, Japan, Singapore, England, France, Germany, Italy and Spain.
Affected count
US: 70 Units; Ex-US 76 Units
Manufactured in
1001 US Route 202, Raritan, NJ, United States
Products
CELLTRACKS ANALYZER II (Software Version 2.5.1 & Below. Product Usage: The CELLTRACKS AUTOPREP System is a general purpose laboratory instrument used with immunomagnetic reagents that capture and enrich target cells, and labeling reagents that differentiate cells in whole blood. Cell analyzers such as the CellTracks Analyzer II, CellSpotter System, flow cytometers or microscopes may be used for cell identification and enumeration. The system is fit for in vitro diagnostic use. In the CellTracks Analyzer II software 2.5.1 release notes Pub. No. J40136EN issued 2012-03-26 page 1, the document states that the edit patient sample information screen now includes a patient information section, a new feature unique to software 2.5.1. This section provides a drop down for patient race and cancer type. There are also separate fields for first and last name for both the physician and patient. There is no mention of the drop down patient ID box on the right side of the user interface which auto-populates previous sample data.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1950-2013Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.