FDA-Device2018-06-06Class II

Sternal Valve Retractors With Atrial and Aortic Rakes; Models SRCL-1, SRCL-AR, SRLP-1, SRLP-AR, ATR-1, ATR-AR, CSR-1 and CSR-AR. For use in cardiac surgery.

Genesee BioMedical, Inc.
Hazard

Failure in a retractor blade due to corrosion. In the area of the blade where the bearing/button meets the blade body, corrosion can form, weakening the welded joint and may cause the bearing/button to break.

Sold states
Worldwide distribution. US nationwide, South Korea, Australia, COSTA RICA, Taiwan, New Zealand, Italy, Canada, Netherlands, Norway, Sweden, Switzerland, Israel, Japan, Egypt, South Africa, United Kingdom
Affected count
444 units
Manufactured in
700 W Mississippi Ave Unit D5, N/A, Denver, CO, United States
Products
Sternal Valve Retractors With Atrial and Aortic Rakes; Models SRCL-1, SRCL-AR, SRLP-1, SRLP-AR, ATR-1, ATR-AR, CSR-1 and CSR-AR. For use in cardiac surgery.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1953-2018

Don't want to check this manually?