FDA-Device2021-07-07Class I
LeadCare Plus Blood Lead Test Kit Catalog Number 82-0004
Magellan Diagnostics, Inc.
Hazard
Magellan has received reports that control tests of either the Low-Control and/or the High-Control generated a low result. The issue may cause the underestimation of blood lead levels in patient samples.
Sold states
Worldwide distribution. US nationwide including Puerto Rico; Canada, Vietnam, India, Italy, Peru, New Zealand, Australia, Hong Kong, and Philippines.
Affected count
320 kits (96 tests/kit=30,720 tests)
Manufactured in
101 Billerica Ave Bldg 4, N/A, North Billerica, MA, United States
Products
LeadCare Plus Blood Lead Test Kit Catalog Number 82-0004
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1953-2021Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief