FDA-Device2012-07-18Class II
CSI Stealth 360-degree Orbital Atherectomy Device (OAD), Part Number 70056-11, Model Number PRD-SC30-125S. Sterilized with Ethylene Oxide. The Stealth 360-degree Orbital Atherectomy Device (OAD) is part of the Stealth 360-degree Orbital PAD system. The Stealth 360-degree Orbital PAD system is a percutaneous orbital atherectomy system indicated for use as therapy in patients with occlusive atherosclerotic disease in peripheral arteries and who are acceptable candidates for percutaneous trans
Cardiovascular Systems, Inc.
Hazard
CSI has initiated a recall on specific lot numbers of Stealth 360-degree Orbital Atherectomy devices which have the potential for an insufficient crown solder bond that may result in the crown becoming detached from the driveshaft.
Sold states
Nationwide distribution including CA, FL, IL, MI, MN, PA, and TX.
Affected count
27
Manufactured in
651 Campus Dr, Saint Paul, MN, United States
Products
CSI Stealth 360-degree Orbital Atherectomy Device (OAD), Part Number 70056-11, Model Number PRD-SC30-125S. Sterilized with Ethylene Oxide. The Stealth 360-degree Orbital Atherectomy Device (OAD) is part of the Stealth 360-degree Orbital PAD system. The Stealth 360-degree Orbital PAD system is a percutaneous orbital atherectomy system indicated for use as therapy in patients with occlusive atherosclerotic disease in peripheral arteries and who are acceptable candidates for percutaneous transluminal atherectomy.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1959-2012Don't want to check this manually?