FDA-Device2021-07-07Class II
Dragonfly Laryngeal Surface Electrode, ET Tube Size (mm): 7.5-10.0, Channel 1, Reference Numbers LSE4246MCA075 and LSE500M. for continuous EMG monitoring of the larynx during surgical procedures.
Spes Medica
Hazard
The supplier informed them of possible deviations of the parameters/processes defined for ethylene oxide sterilization.
Sold states
Distribution in the following US states: CA, MA, and SC.
Affected count
20 boxes
Manufactured in
Via Buccari 21, N/A, Genoa, N/A, Italy
Products
Dragonfly Laryngeal Surface Electrode, ET Tube Size (mm): 7.5-10.0, Channel 1, Reference Numbers LSE4246MCA075 and LSE500M. for continuous EMG monitoring of the larynx during surgical procedures.