FDA-Device2016-06-22Class II
Vaxcel PICC with PASV IR-145 Kit under the following labels: 1) 4F, UPN M001454580, Rx ONLY, 2) 5F, UPN M001454630, Rx ONLY & 3) 6F, UPN M001454780, Rx ONLY The Vaxcel PICC with PASV is indicated for use in establishing peripheral access to the central venous system for administration of fluids including, but not limited to, hydration agents, antibiotics, chemotherapy, analgesics, nutritional therapy, and blood products. It is also indicated for blood specimen withdrawal. This product is intend
Navilyst Medical, Inc., an AngioDyamics Company
Hazard
PICC catheters that contain valve(s) manufactured prior to July 15, 2014 lack a manufacturing specification change that reduces the incidence of hemolysis during blood sampling through the PICC catheter.
Sold states
Worldwide Distribution US (Nationwide) Internationally to AU, GR, SE, SA, IN, CA, GB, BR, BE, NL, ES, KR, HK, AE, IT and ES.
Affected count
2,260 units
Manufactured in
10 Glens Falls Tech Park, Glens Falls, NY, United States
Products
Vaxcel PICC with PASV IR-145 Kit under the following labels: 1) 4F, UPN M001454580, Rx ONLY, 2) 5F, UPN M001454630, Rx ONLY & 3) 6F, UPN M001454780, Rx ONLY The Vaxcel PICC with PASV is indicated for use in establishing peripheral access to the central venous system for administration of fluids including, but not limited to, hydration agents, antibiotics, chemotherapy, analgesics, nutritional therapy, and blood products. It is also indicated for blood specimen withdrawal. This product is intended for central venous access in adults, children and infants who require intravenous (IV) therapy.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1968-2016Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief