FDA-Device2015-07-08Class II

Infinite 200 PRO in vitro diagnostic. Product Usage: The Tecan Infinite 200 PRO is a multifunctional microplate reader with injector option. The Infinite 200 PRO provides high performance for the vast majority of todays microplate applications and research and is robotic compatible. The Infinite 200 PRO has been designed as a general purpose laboratory instrument for professional use, supporting common 6 to 384 well microplates conforming to the ANS/SBS standards.

Tecan US, Inc.
Hazard

A firmware bug causes incorrect values to be displayed for specific measured luminescence values.

Sold states
US Nationwide Distribution in the states of AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MO, NE NH, NJ, NM, NY, NC, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, including DC
Affected count
868 in total
Manufactured in
9401 Globe Center Drive, Suite 140, Morrisville, NC, United States
Products
Infinite 200 PRO in vitro diagnostic. Product Usage: The Tecan Infinite 200 PRO is a multifunctional microplate reader with injector option. The Infinite 200 PRO provides high performance for the vast majority of todays microplate applications and research and is robotic compatible. The Infinite 200 PRO has been designed as a general purpose laboratory instrument for professional use, supporting common 6 to 384 well microplates conforming to the ANS/SBS standards.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1970-2015