FDA-Device2016-06-22Class II
Xcela Hybrid PICC with PASV Valve Technology, MST-70 Kit, 6F-55cm, UPN H965952420 & UPN H965952460, Rx ONLY The Xcela PICC with PASV, Xcela Hybrid PICC with PASV, BioFlo PICC with PASV and BioFlo Hybrid PICC with PASV are indicated for short or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications and nutrients; the sampling of blood; and for power injection of contrast media. Non-valve lumens ar
Navilyst Medical, Inc., an AngioDyamics Company
Hazard
PICC catheters that contain valve(s) manufactured prior to July 15, 2014 lack a manufacturing specification change that reduces the incidence of hemolysis during blood sampling through the PICC catheter.
Sold states
Worldwide Distribution US (Nationwide) Internationally to AU, GR, SE, SA, IN, CA, GB, BR, BE, NL, ES, KR, HK, AE, IT and ES.
Affected count
157 units
Manufactured in
10 Glens Falls Tech Park, Glens Falls, NY, United States
Products
Xcela Hybrid PICC with PASV Valve Technology, MST-70 Kit, 6F-55cm, UPN H965952420 & UPN H965952460, Rx ONLY The Xcela PICC with PASV, Xcela Hybrid PICC with PASV, BioFlo PICC with PASV and BioFlo Hybrid PICC with PASV are indicated for short or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications and nutrients; the sampling of blood; and for power injection of contrast media. Non-valve lumens are indicated for central venous pressure monitoring
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1979-2016Don't want to check this manually?