FDA-Device2021-07-07Class II
Physician Order Form (PL-15400-00, Nov. 2018) distributed to end user used with the following: Prometra Programmable Infusion Pump System: MODEL/CATALOG #(s): Catalog No. US Model GTIN-14 11827 Prometra Programmable 20mL Pump 00810335020082 13827 Prometra II Programmable 20mLPump 00810335020228 16827 Prometra II Programmable 40mLPump 00810335020242
Flowonix Medical Inc
Hazard
Physician Order Form (PL-15400-00 Nov 2018) distributed is discontinued due to the text 'Refill the pump and program refill with the same medication removed prior to MRI' may be misinterpreted as 'reuse the same drug solution extracted from the pump prior to the MRI procedure', creating a possible risk of infection
Sold states
US Nationwide distribution.
Affected count
N/A
Manufactured in
120 Forbes Blvd Ste 170, N/A, Mansfield, MA, United States
Products
Physician Order Form (PL-15400-00, Nov. 2018) distributed to end user used with the following: Prometra Programmable Infusion Pump System: MODEL/CATALOG #(s): Catalog No. US Model GTIN-14 11827 Prometra Programmable 20mL Pump 00810335020082 13827 Prometra II Programmable 20mLPump 00810335020228 16827 Prometra II Programmable 40mLPump 00810335020242
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1996-2021Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief