FDA-Device2012-07-25Class I

CareFusion EnVe Ventilator Designed for use on patients who require respiratory support or mechanical ventilation and weigh a minimum of 5 kg (11 lbs). It is suitable for service in hospital and transport environments as a source of continuous or intermittent positive pressure ventilatory support, delivered invasively or non-invasively. It is not intended for homecare use.

CareFusion 203, Inc.
Hazard

CareFusion identified that the EnVe Ventilator may not hold the set Positive End Expiratory Pressure (PEEP) value either intermittently or continuously or it may alarm for a Patient Circuit Fault. The unit will activate either a Low PEEP or Patient Circuit Fault Alarm both audibly and Visually to alert the healthcare professional. Ventilation delivery to the patient may be compromised.

Sold states
Nationwide Distribution including AZ, CA, CO, FL, IL, KY, ME, MI, MO, MN, MT, ND, NJ, NM, NV, NY, OK, PA, SC, SD, TX , UT, WI, and WV.
Affected count
266
Manufactured in
17400 Medina Rd Ste 100, Minneapolis, MN, United States
Products
CareFusion EnVe Ventilator Designed for use on patients who require respiratory support or mechanical ventilation and weigh a minimum of 5 kg (11 lbs). It is suitable for service in hospital and transport environments as a source of continuous or intermittent positive pressure ventilatory support, delivered invasively or non-invasively. It is not intended for homecare use.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2006-2012