FDA-Device2013-08-28Class II
65597201333 Femur CR POR FEM SURF HDN HATCP CML, Rx, Sterile; 65597201334 Femur CR POR FEM SURF HDN HATCP CMR, Rx, Sterile; 65597201433 Femur CR POR FEM SURF HDN HATCP DML, Rx, Sterile; 65597201434 Femur CR POR FEM SURF HDN HATCP DMR, Rx, Sterile; 65597201533 Femur CR POR FEM SURF HDN HATCP EML, Rx, Sterile; 65597201534 Femur CR POR FEM SURF HDN HATCP EMR, Rx, Sterile; Used in total knee arthroplasty and indicated for patients with severe pain.
Zimmer, Inc.
Hazard
Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs, Articular Surface, Patellae, instruments and provisionals combined with standard components. A March 2012 recall reiterated the proper combination labeling, but the mis match problem continues. Zimmer is removing the d
Sold states
Worldwide Distribution-USA (nationwide) and the country of Japan.
Affected count
192,355 all devices
Manufactured in
345 E Main St, N/A, Warsaw, IN, United States
Products
65597201333 Femur CR POR FEM SURF HDN HATCP CML, Rx, Sterile; 65597201334 Femur CR POR FEM SURF HDN HATCP CMR, Rx, Sterile; 65597201433 Femur CR POR FEM SURF HDN HATCP DML, Rx, Sterile; 65597201434 Femur CR POR FEM SURF HDN HATCP DMR, Rx, Sterile; 65597201533 Femur CR POR FEM SURF HDN HATCP EML, Rx, Sterile; 65597201534 Femur CR POR FEM SURF HDN HATCP EMR, Rx, Sterile; Used in total knee arthroplasty and indicated for patients with severe pain.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2013-2013Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief