FDA-Device2013-08-28Class II
00595204110 Articular Surface XLPE CR ART SURF 5-6/STRGRN 10, Rx, Sterile; 00595204112 Articular Surface XLPE CR ART SURF 5-6/STRGRN 12, Rx, Sterile; 00595204114 Articular Surface XLPE CR ART SURF 5-6/STRGRN 14, Rx, Sterile; 00595204117 Articular Surface XLPE CR ART SURF 5-6/STRGRN 17, Rx, Sterile; 00595204120 Articular Surface XLPE CR ART SURF 5-6/STRGRN 20, Rx, Sterile; Used in total knee arthroplasty and indicated for patients with severe pain.
Zimmer, Inc.
Hazard
Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs, Articular Surface, Patellae, instruments and provisionals combined with standard components. A March 2012 recall reiterated the proper combination labeling, but the mis match problem continues. Zimmer is removing the d
Sold states
Worldwide Distribution-USA (nationwide) and the country of Japan.
Affected count
192,355 all devices
Manufactured in
345 E Main St, N/A, Warsaw, IN, United States
Products
00595204110 Articular Surface XLPE CR ART SURF 5-6/STRGRN 10, Rx, Sterile; 00595204112 Articular Surface XLPE CR ART SURF 5-6/STRGRN 12, Rx, Sterile; 00595204114 Articular Surface XLPE CR ART SURF 5-6/STRGRN 14, Rx, Sterile; 00595204117 Articular Surface XLPE CR ART SURF 5-6/STRGRN 17, Rx, Sterile; 00595204120 Articular Surface XLPE CR ART SURF 5-6/STRGRN 20, Rx, Sterile; Used in total knee arthroplasty and indicated for patients with severe pain.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2028-2013Don't want to check this manually?