FDA-Device2021-07-14Class II

FREELITE Human Kappa Free Kit for use on the Roche cobas c Systems, Kappa Freie Leichtkette (Human) kit for Roche cobas c Systems, REF: LK016.CB, IVD CE, UDI: 05051700012096

The Binding Site Group, Ltd.
Hazard

The high and low level "Quality Controls" demonstrate a positive bias of approximately 10% compared to assigned values.

Sold states
Worldwide distribution - US Nationwide distribution in the states of CA, GA, FL, MA, MO, MI, WA, NY, PA, MN, AL, MD, OK, NE, IN, ID, OH, TX, HI, DC and the countries of Canada, Australia, Austria, Belgium, Bulgaria, Canada, Cyprus, Czech Republic, Denmark, Estonia, France, Germany, Italy, Luxembourg, Netherlands, Norway, Poland, Portugal, Romania, Slovenia, Spain, Switzerland, United Kingdom, Argentina, Egypt, GUADELOUPE, HONG KONG, India, Iran, Israel, Korea, Lebanon, Malaysia, Morocco, Oman, Singapore, Taiwan, Tunisia, Vietnam.
Manufactured in
8 Calthorpe Road, Birmingham, United Kingdom
Products
FREELITE Human Kappa Free Kit for use on the Roche cobas c Systems, Kappa Freie Leichtkette (Human) kit for Roche cobas c Systems, REF: LK016.CB, IVD CE, UDI: 05051700012096

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2030-2021

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief