FDA-Device2012-08-01Class I

OEC 6800 Miniview, Image-intensified fluoroscopic x-ray system used during diagnostic, surgical and interventional procedures. Designed to provide the physician with general fluoroscopic visualization of the patient including but not limited to surgical orthopedic and extremity imagine. The device is not intended for whole-body pediatric imaging.

GE OEC Medical Systems, Inc
Hazard

Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. The OEC 6800 Miniview was recalled in November 2006 due to several intermittent potential safety issues including navigation tracking inaccuracy, system Lock-up, incorrect image display, incorrect dose area calculated air kerma in "film mo

Sold states
Worldwide Distribution
Affected count
1,933 units
Manufactured in
384 Wright Brothers Dr, Salt Lake City, UT, United States
Products
OEC 6800 Miniview, Image-intensified fluoroscopic x-ray system used during diagnostic, surgical and interventional procedures. Designed to provide the physician with general fluoroscopic visualization of the patient including but not limited to surgical orthopedic and extremity imagine. The device is not intended for whole-body pediatric imaging.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2040-2012