FDA-Device2012-08-01Class II
Universal Spinal System Lamina Hooks Synthes Medium TI Lamina Hook - Right 498.320 Non sterile LOT 6876785 Noncervical spinal fixation devices intended for posterior pedicle screw fixation, posterior hook fixation, or anterolateral fixation. Limited to skeletally mature patients with the exception of the small stature USS, which includes small stature and pediatric patients. These devices are indicated as an adjunct to fusion for all of these indications: degenerative disc disease, spond
Synthes (USA) Products LLC
Hazard
USS Lamina Hooks, Medium, Right are being recalled because the hooks are etched with the wrong part number - 498.321 (Lamina Hook, Medium, Left). The Lamina Hooks should be etched with part number 498.320 (Lamina Hook, Medium, right).
Sold states
Nationwide Distribution including GA, PA, SC, TX, and UT and Canada.
Affected count
12 units
Manufactured in
1301 Goshen Pkwy, West Chester, PA, United States
Products
Universal Spinal System Lamina Hooks Synthes Medium TI Lamina Hook - Right 498.320 Non sterile LOT 6876785 Noncervical spinal fixation devices intended for posterior pedicle screw fixation, posterior hook fixation, or anterolateral fixation. Limited to skeletally mature patients with the exception of the small stature USS, which includes small stature and pediatric patients. These devices are indicated as an adjunct to fusion for all of these indications: degenerative disc disease, spondylolisthesis, trauma, deformities or curvatures, tumor, stenosis and failed previous fusion.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2054-2012Don't want to check this manually?