FDA-Device2021-07-21Class II
Discovery MR750w 3.0T - Product Usage: is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times.
GE Healthcare, LLC
Hazard
GE Healthcare has recently become aware of an issue on the affected products listed below where the "Patient Orientation" button may inadvertently be clicked when intending to click on the "Save RX" button. This changes the prescribed patient orientation on the system prior to running the initial 3-Plane Localizer Scan (See Figure 1 for reference to buttons.) Selecting and saving a patient orientation that does not match the patient's actual position may result in incorrectly annotated and/or flipped images.
Sold states
Worldwide distributions.
Affected count
585 devices
Manufactured in
3000 N Grandview Blvd, Waukesha, WI, United States
Products
Discovery MR750w 3.0T - Product Usage: is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2055-2021Don't want to check this manually?