FDA-Device2021-07-21Class II

SIGNA PET/MR, Tomographic Imager Combining Emission Computed Tomography With Nuclear Magnetic Resonance

GE Healthcare, LLC
Hazard

GE Healthcare has recently become aware of an issue on the affected products listed below where the "Patient Orientation" button may inadvertently be clicked when intending to click on the "Save RX" button. This changes the prescribed patient orientation on the system prior to running the initial 3-Plane Localizer Scan (See Figure 1 for reference to buttons.) Selecting and saving a patient orientation that does not match the patient's actual position may result in incorrectly annotated and/or flipped images.

Sold states
Worldwide distributions.
Affected count
71 devices
Manufactured in
3000 N Grandview Blvd, Waukesha, WI, United States
Products
SIGNA PET/MR, Tomographic Imager Combining Emission Computed Tomography With Nuclear Magnetic Resonance

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2057-2021

Don't want to check this manually?