FDA-Device2017-05-17Class II

Medi-Therm Hyper/Hypothermia System, MTA7900

Stryker Medical Division of Stryker Corporation
Hazard

The water may reheat too quickly if the unit is in "Automatic" mode, switched to "Manual" mode to view water temperature, and then switched back to "Automatic" mode.

Sold states
USA (nationwide) Distribution and Internationally to Colombia.
Affected count
119
Manufactured in
3800 E Centre Ave, N/A, Portage, MI, United States
Products
Medi-Therm Hyper/Hypothermia System, MTA7900

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2058-2017

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief