Medi-Therm Hyper/Hypothermia System, MTA7900
Stryker Medical Division of Stryker Corporation
Hazard
The water may reheat too quickly if the unit is in "Automatic" mode, switched to "Manual" mode to view water temperature, and then switched back to "Automatic" mode.
Sold states
USA (nationwide) Distribution and Internationally to Colombia.
Affected count
119
Manufactured in
3800 E Centre Ave, N/A, Portage, MI, United States
Products
Medi-Therm Hyper/Hypothermia System, MTA7900